Technical Data

• compact turn-key tunable Ti:sapphire femtosecond laser

laser pulse width:

< 100 fs

repetition frequency:

80 / 90 MHz

mean laser output:

0 ... 1.5 W

wavelength range:

720 ... 920 nm

• full-frame scanning, region-of-interest (ROI) line scanning,
single-point illumination (spot scan)

• typical scan range:

350 × 350 μm (horizontal)

 

200 μm (vertical)

• spatial resolution:

< 1 μm (horizontal)

 

< 2 μm (vertical)

• temporal resolution (FLIM):

250 ps

• focusing optics:

magnification 40×

 

numerical aperture (NA) 1.3

• video adapter for visualization with CCD-camera

• control and image processing software (JenLab Image)

• operating temperature:

15 ... 35 °C (59 ... 95 °F)

• relative humidity:

5 ... 95 % (non-condensing)

• power requirements:

230 VAC (50 Hz) or
115 VAC (60 Hz)

• CE certified

 

 

 

System dimensions

• Workstation:

1200 × 780 × 1100 mm3
180 kg

• Scan module:

615 × 255 × 210 mm3
16 kg

• Control unit:

450 × 460 × 190 mm3
12 kg

• Ti:sapph. laser:

600 × 370 × 180 mm3
42 kg (laser head)
41 kg (power supply)

• Chiller:

270 × 200 × 380 mm3
20 kg

The overall system needs a minimum of 8 m2. Air-condition is recommended.
For operation reduce ambient light.

Notes: This specification is subject to change without notice.
DermaInspect® is not intended as a primary means of diagnosis.